American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
- Products
-
- AMERICAN CONTRACT SYSTEMS
- Where it was sold
-
US distribution to states of: IL and LA
States: Illinois, Louisiana
- Barcodes
0019107223637100191072238771- Units affected
- 50 units
- About this recall
- AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ROBOT HYST PACK, MODEL/ITEM NUMBER GRRB10E; 2) VAG HYST PACK, MODEL/ITEM NUMBER SPVH62P.