Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Products
-
- MEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Nationwide
- Barcodes
10195327191887101953272877641019532743211910198459272127101953271001311019532738137010195327100148- Units affected
- 1432 units
- About this recall
- MEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483C; 4) CV DRAPING/BASIN 2PART PACK, Medline Kit Number/SKU DYNJ89414; 5) CVG BASIN PACK SC, Medline Kit Number/SKU DYNJ38052G; 6) T/A BASIN SET, Medline Kit Number/SKU DYNJ47941F; 7) VALVE BASIN PACK SC, Medline Kit Number/SKU DYNJ38054G.