Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Products
-
- MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Nationwide
- Barcodes
101934897665541019532768466210198459518973- Units affected
- 342 units
- About this recall
- MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medline Kit Number/SKU CDS983800K.