Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Products
  • MEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK
Where it was sold

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Nationwide

Barcodes
108899421825891019845967604810195327205980101953274088861019348935250410195327446635101953271910921019845949582310198459572753101953273506351019532763765110198459485565101934895558201019532718822110889942572595
Units affected
24388 units
About this recall
MEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK, Medline Kit Number/SKU DYNJ47625B; 2) GYN VAGINAL, Medline Kit Number/SKU DYNJ912219; 3) KIT MAJ VAG SET UP, Medline Kit Number/SKU DYNJ907033B; 4) MINOR VAG PACK, Medline Kit Number/SKU DYNJ68663B; 5) OB DELIVERY PACK, Medline Kit Number/SKU DYNJ14099F; 6) OB DELIVERY PACK, Medline Kit Number/SKU DYNJ14099G; 7) OB PACK, Medline Kit Number/SKU DYNJ03387B; 8) OB PACK, Medline Kit Number/SKU DYNJ03387C; 9) OB PACK, Medline Kit Number/SKU DYNJ53543B; 10) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645D; 11) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645F; 12) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645G; 13) VAGINAL DELIVERY PACK, Medline Kit Number/SKU DYNJ04454K; 14) VAGINAL DELIVERY PACK, Medline Kit Number/SKU DYNJ04454L; 15) VAGINAL PACK, Medline Kit Number/SKU DYNJ01830G.