Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Products
-
- MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Nationwide
- Barcodes
101934896404581019532719347810195327451684101953270249631019532716538310193489981728101934899511271019532705745910195327480844101984595695621019348998508510193489827798101953273993991019845924704010195327635657101984595721801019845924649410195327568689101984594771881019532742529610198459405792101934899473661019348988837910198459107009101953274141841019532761988610198459410055101984596370701019532703031510198459287152101984593930681019845908221410198459255618101934894997281088827776362310888277783683101934897535611019348949068810195327113780101934897502491019845943636910193489759389101953270480821019845943637610193489733747101934899122411019845937447010193489932164101953271116701019532747030210193489906059101984596175601019532753291810195327095550101934893420481019532766921810195327597306108899420237381088827794206610889942666287101953270794511019348998109410889942435951108843896344931088994263009710193489432701101934899834561019348998245910195327251079101953275761341088994271693710193489915693101934899631681019532705695710195327221164101953273201881019348992096310195327042905101984596254731088994240837510193489396744108899421139651019532750894410193489950830- Units affected
- 37324 units
- About this recall
- MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, LAVH NEW BRAUNFELS,TVH PACK, etc (see recall documents for a full list of affected products)