Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Products
  • MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY
Where it was sold

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Nationwide

Barcodes
101934896404581019532719347810195327451684101953270249631019532716538310193489981728101934899511271019532705745910195327480844101984595695621019348998508510193489827798101953273993991019845924704010195327635657101984595721801019845924649410195327568689101984594771881019532742529610198459405792101934899473661019348988837910198459107009101953274141841019532761988610198459410055101984596370701019532703031510198459287152101984593930681019845908221410198459255618101934894997281088827776362310888277783683101934897535611019348949068810195327113780101934897502491019845943636910193489759389101953270480821019845943637610193489733747101934899122411019845937447010193489932164101953271116701019532747030210193489906059101984596175601019532753291810195327095550101934893420481019532766921810195327597306108899420237381088827794206610889942666287101953270794511019348998109410889942435951108843896344931088994263009710193489432701101934899834561019348998245910195327251079101953275761341088994271693710193489915693101934899631681019532705695710195327221164101953273201881019348992096310195327042905101984596254731088994240837510193489396744108899421139651019532750894410193489950830
Units affected
37324 units
About this recall
MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, LAVH NEW BRAUNFELS,TVH PACK, etc (see recall documents for a full list of affected products)