Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Products
  • MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS
Where it was sold

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Nationwide

Barcodes
1019348998066010195327684761101984590869911019845951841610198459658358101953273701211019348928449210195327325695
Units affected
2340 units
About this recall
MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775K; 4) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775L; 5) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775M; 6) TONSIL PACK, Medline Kit Number/SKU DYNJ84480; 7) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080; 8) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080A.