Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: 1) CRANI

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Products
  • MEDLINE Medical Procedure Kits labeled as: 1) CRANI
Where it was sold

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Nationwide

Barcodes
101934899738461019532750707710198459043499101984591297731019845923778210198459600197108899428864181088994262406510193489430615101953276061691019348933623810195327604967
Units affected
3521 units
About this recall
MEDLINE Medical Procedure Kits labeled as: 1) CRANI, Medline Kit Number/SKU DYNJ908097; 2) CRANI, Medline Kit Number/SKU DYNJ908097A; 3) CRANI, Medline Kit Number/SKU DYNJ908097B; 4) CRANI, Medline Kit Number/SKU DYNJ908097C; 5) CRANI, Medline Kit Number/SKU DYNJ908097D; 6) DR TUCKER ANT TOTAL HIP, Medline Kit Number/SKU DYNJ912029; 7) JEWISH CRANI PACK, Medline Kit Number/SKU DYNJ55637B; 8) L SCOPE PACK, Medline Kit Number/SKU DYNJ43930A; 9) LAMI PACK, Medline Kit Number/SKU DYNJ67767; 10) LAVH NEW BRAUNFELS, Medline Kit Number/SKU DYNJ61136F; 11) SOI KNEE ARTHROSCOPY, Medline Kit Number/SKU DYNJV0172Q; 12) T AND A PACK, Medline Kit Number/SKU DYNJ86229.