Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ICU Medical Inc. recalls Smiths Medical CADD" Medication Cassette Reservoir with flow stop

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
Products
  • Smiths Medical CADD" Medication Cassette Reservoir with flow stop
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
10610586027215
Units affected
211750 units
About this recall
Smiths Medical CADD" Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 50mL, Product Code 21-7301-24