ICU Medical Inc. recalls Smiths Medical CADD" Medication Cassette Reservoir with flow stop
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- Products
-
- Smiths Medical CADD" Medication Cassette Reservoir with flow stop
- Company
- ICU Medical Inc.
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
10610586027215- Units affected
- 211750 units
- About this recall
- Smiths Medical CADD" Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 50mL, Product Code 21-7301-24