Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
- Products
-
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, and TX. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
States: California, Florida, Georgia, Indiana, Kansas, Maine, New Hampshire, New Jersey, Oklahoma, Pennsylvania, Texas
- Barcodes
15099590732103- Units affected
- 183 devices
- About this recall
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.