Beckman Coulter, Inc. recalls Discrete photometric chemistry analyzer for clinical use.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
- Products
-
- Discrete photometric chemistry analyzer for clinical use.
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.
States: Alabama, California, Illinois, Louisiana, Michigan, Montana, Nebraska, New York, Texas
- Units affected
- 59 units
- About this recall
- Discrete photometric chemistry analyzer for clinical use.