Baxter Healthcare Corporation recalls Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit)

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Products
  • Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit)
Where it was sold

Worldwide distribution - US Nationwide and the country of Canada.

Nationwide

Barcodes
10887761985015
Units affected
0
About this recall
Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device