Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Baxter Healthcare Corporation recalls Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit)

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Products
  • Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit)
Where it was sold

Worldwide distribution - US Nationwide and the country of Canada.

Nationwide

Barcodes
00887761981492
Units affected
16187 units
About this recall
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device