Baxter Healthcare Corporation recalls Hillrom VOLARA System P.CIRCUIT 5KIT AC
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
- Products
-
- Hillrom VOLARA System P.CIRCUIT 5KIT AC
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Canada.
Nationwide
- Barcodes
10887761981499- Units affected
- 13881 units
- About this recall
- Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device