Stryker Neurovascular recalls INZONE DETACHMENT SYSTEM

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
Products
  • INZONE DETACHMENT SYSTEM
Where it was sold

Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUSTRIA, BELGIUM, BRASIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, KOREA, LATVIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH PACIFIC, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.

Nationwide

Barcodes
04546540697950
Units affected
44,937
About this recall
INZONE DETACHMENT SYSTEM, REF: M00345100950