Orthofix U.S. LLC recalls PILLAR SA Ti

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A 37 mm implant width intervertebral body fusion device has a labeling discrepancy with incorrect width of 33 mm stated on the outer label side flap and inner label side flap information box. The 37 mm width on the front labels is correct.
Products
  • PILLAR SA Ti
Where it was sold

US Nationwide distribution in the states of CO, CA.

States: California, Colorado

Barcodes
18257200161010
Units affected
3
About this recall
PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP