Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
- Products
-
- PORTEX Bivona Tracheostomy Silicone Tube
- Company
- Smiths Medical ASD Inc.
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
15021312005721150213120057381502131200574515021312005752150213120057691502131200577615021312005783- About this recall
- PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM UNCUFFED , Product Code/List Number/Item Code 60P025; b) PAED. TRACHEOSTOMY TUBE 3.0MM UNCUFFED , Product Code/List Number/Item Code 60P030; c) PAED. TRACHEOSTOMY TUBE 3.5MM UNCUFFED , Product Code/List Number/Item Code 60P035; d) PAED. TRACHEOSTOMY TUBE 4.0MM UNCUFFED , Product Code/List Number/Item Code 60P040; e) PAED. TRACHEOSTOMY TUBE 4.5MM UNCUFFED , Product Code/List Number/Item Code 60P045; f) PAED. TRACHEOSTOMY TUBE 5.0MM UNCUFFED , Product Code/List Number/Item Code 60P050; g) PAED. TRACHEOSTOMY TUBE 5.5MM UNCUFFED , Product Code/List Number/Item Code 60P055