Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
- Products
-
- PORTEX Bivona Tracheostomy Silicone Tube
- Company
- Smiths Medical ASD Inc.
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
10351688518774103516885187811035168851879810351688518804- About this recall
- PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND STRAIGHT FLANGE , Product Code/List Number/Item Code 67NFPS25; b) NEONATAL TRACHEOSTOMY TUBE 3.0mm TTS FLEXTEND STRAIGHT FLANGE , Product Code/List Number/Item Code 67NFPS30; c) NEONATAL TRACHEOSTOMY TUBE 3.5mm TTS FLEXTEND STRAIGHT FLANGE , Product Code/List Number/Item Code 67NFPS35; d) NEONATAL TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND STRAIGHT FLANGE , Product Code/List Number/Item Code 67NFPS40