Medline Industries, LP recalls Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
- Products
-
- Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline
- Company
- Medline Industries, LP
- Where it was sold
-
Domestic US distribution nationwide. International distribution pending. No international distribution.
Nationwide
- Units affected
- 7913 eaches
- About this recall