Medline Industries, LP recalls Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during di
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
- Products
-
- Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during di
- Company
- Medline Industries, LP
- Where it was sold
-
Domestic US distribution nationwide. International distribution pending. No international distribution.
Nationwide
- Units affected
- 7913 eaches
- About this recall
- Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH