DKK Dai-Ichi Shomei Co., Ltd. recalls Aurora 2 Series
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.
- Products
-
- Aurora 2 Series
- Where it was sold
-
Domestic: Nationwide Distribution International: Australia, Canada, Egypt, Hong Kong, Indonesia, Japan, Jordan, Netherlands, New Zealand, Philippines, Singapore, South America, South Korea, Thailand, UAE.
Nationwide
- Units affected
- 822 units
- About this recall
- Aurora 2 Series, Aurora 7 lighthead; Models: AUR7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.