Masimo Corporation recalls Radius VSM Disposable NIBP Cuff
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
- Products
-
- Radius VSM Disposable NIBP Cuff
- Company
- Masimo Corporation
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of OK, CA , MD , MA , NY , UT , NC , TN , MN , TX , OH , PA , VA and the countries of Italy, Latvia, Iraq, Saudi Arabia, Germany, Spain, Turkey, France, Australia, Malta, Oman, United Arab Emirates, Martinique, Netherlands, Cyprus, Kuwait, Czech Republic, United Kingdom, Croatia, Switzerland, Belgium, Singapore, Romania, Colombia, Hong Kong, Israel, Finland, Poland, New Zealand, Tunisia, Chile, Algeria, Canada, Denmark, Palestinian Territory, Qatar, Greece, Paraguay.
States: California, Massachusetts, Maryland, Minnesota, North Carolina, New York, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, Virginia
- Units affected
- 1,028
- About this recall
- Radius VSM Disposable NIBP Cuff, REF: 4825, 4826