Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ICU Medical, Inc. recalls ICU Medical Plum Duo Infusion Pump

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Products
  • ICU Medical Plum Duo Infusion Pump
Where it was sold

Worldwide distribution - US Nationwide and the country of Canada.

Nationwide

Units affected
12525 units
About this recall
ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump