Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Daavlin Distributing Company recalls Photera 600 3-D

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Products
  • Photera 600 3-D
Where it was sold

US Nationwide distribution.

Nationwide

Units affected
212 units
About this recall
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders