Kent Imaging, Inc. recalls SnapshotGLO Wound Imaging Device
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
- Products
-
- SnapshotGLO Wound Imaging Device
- Company
- Kent Imaging, Inc.
- Where it was sold
-
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
States: Arkansas, Arizona, California, District of Columbia, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Nevada, Pennsylvania, Tennessee, Texas, Utah
- Units affected
- 77
- About this recall
- SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.