Kent Imaging, Inc. recalls SnapshotGLO Wound Imaging Device

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Products
  • SnapshotGLO Wound Imaging Device
Where it was sold

US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.

States: Arkansas, Arizona, California, District of Columbia, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Nevada, Pennsylvania, Tennessee, Texas, Utah

Units affected
77
About this recall
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.