Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Alphatec Spine, Inc. recalls SafeOp Patient Interface

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.
Products
  • SafeOp Patient Interface
Where it was sold

US Nationwide distribution in the states of TX, MO, IN.

States: Indiana, Missouri, Texas

Barcodes
00190376517506
Units affected
3
About this recall
SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System