Alphatec Spine, Inc. recalls SafeOp Patient Interface
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.
- Products
-
- SafeOp Patient Interface
- Company
- Alphatec Spine, Inc.
- Where it was sold
-
US Nationwide distribution in the states of TX, MO, IN.
States: Indiana, Missouri, Texas
- Barcodes
00190376517506- Units affected
- 3
- About this recall
- SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System