Bard Access Systems, Inc. recalls BD Needle Kit for Powered Driver 15mm x 15Ga
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy
- Products
-
- BD Needle Kit for Powered Driver 15mm x 15Ga
- Company
- Bard Access Systems, Inc.
- Where it was sold
-
US: MA, FL, IA, SC, IL, NY, UT, NJ, ID, IN, PA, CA, TX, OH, MO, VA, MD, NC, AL, MN, CO, KY, LA, NV, OK, MS, AZ, MI, GA, AR, KS, TN. OUS: CA, SG
States: Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia
- Units affected
- 80,200
- About this recall
- BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK