Biofire Defense recalls FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Pane
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.
- Products
-
- FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Pane
- Company
- Biofire Defense
- Where it was sold
-
U.S. Nationwide distribution in the states of AL, CA, DC, FL, GA, HI, KY, LA, MA, MD, MN, NC, NH, NV, NY, OH, RI, TN, TX, UT, VA and WA.
States: Alabama, California, District of Columbia, Florida, Georgia, Hawaii, Kentucky, Louisiana, Massachusetts, Maryland, Minnesota, North Carolina, New Hampshire, Nevada, New York, Ohio, Rhode Island, Tennessee, Texas, Utah, Virginia, Washington
- Barcodes
00851458005136- Units affected
- 122 kits (3,660 pouches)
- About this recall
- FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).