Instrumentation Laboratory recalls GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 000
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
- Products
-
- GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 000
- Company
- Instrumentation Laboratory
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
States: Arizona, California, Florida, Georgia, Illinois, Indiana, Kentucky, Minnesota, Missouri, Montana, Ohio, Pennsylvania, Virginia, Wisconsin, West Virginia
- Barcodes
0842695080759908426950807605084269508075370842695080769808426950807759084269508077660842695080781008426950807827- Units affected
- 109 units
- About this recall
- GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A