Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Waldemar Link GmbH & Co. KG (Mfg Site) recalls Endo-Model

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.
Products
  • Endo-Model
Where it was sold

Domestic US distribution to Indiana. OUS distribution to Belarus and Germany.

States: Indiana

Barcodes
04026575432134
Units affected
1 US; 9 OUS
About this recall
Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16