Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

B Braun Medical Inc recalls Infusomat Space Infusion System/ Large Volume Pump

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.
Products
  • Infusomat Space Infusion System/ Large Volume Pump
Where it was sold

US distribution to AL, CA, GA, IL, MI, and TX.

States: Alabama, California, Georgia, Illinois, Michigan, Texas

Units affected
14
About this recall
Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U