Pacific Lasertech, LLC recalls 25-LHP-928
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
- Products
-
- 25-LHP-928
- Company
- Pacific Lasertech, LLC
- Where it was sold
-
US Nationwide Distribution
Nationwide
- Units affected
- 4
- About this recall
- 25-LHP-928, HeNe Laser System, 25 mW