Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Pacific Lasertech, LLC recalls 25-LHP-928

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
Products
  • 25-LHP-928
Where it was sold

US Nationwide Distribution

Nationwide

Units affected
4
About this recall
25-LHP-928, HeNe Laser System, 25 mW