Beckman Coulter Biomedical GmbH recalls PE Centrifuge
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- Products
-
- PE Centrifuge
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.
Nationwide
- Barcodes
15099590673550- About this recall
- PE Centrifuge, REF: B36365, a component of the Power Express