Medline Industries, LP recalls Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
- Products
-
- Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Panama.
Nationwide
- Barcodes
1019734401075210197344010738101973440107211019734401071410197344010677101973440109051019734401092910197344010493101973440104861019734401047910197344010462- Units affected
- 271406 units
- About this recall
- Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).