Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

IMRIS Imaging Inc recalls ORT200

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Products
  • ORT200
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.

Nationwide

Barcodes
00857534006349
Units affected
6 units
About this recall