Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Medline Convenience Kits: 1) DYKM1414

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
Products
  • Medline Convenience Kits: 1) DYKM1414
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
10889942481293101934899934241019348927717310193489311730
Units affected
58037 kits
About this recall
Medline Convenience Kits: 1) DYKM1414, Catalog Number: DYKM1414; 2) DYKM1414, Catalog Number: DYKM1414; 3) DYNDH1462A, Catalog Number: DYNDH1462A; 4) DYNJ38145F, Catalog Number: DYNJ38145F; 5) DYNJ63199A, Catalog Number: DYNJ63199A