Medline Industries, LP recalls Medline Convenience Kits: 1) IVS BLOOD CULTURE KIT
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
- Products
-
- Medline Convenience Kits: 1) IVS BLOOD CULTURE KIT
- Company
- Medline Industries, LP
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
1065316036631710653160385943- Units affected
- 17,750 kits
- About this recall
- Medline Convenience Kits: 1) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920; 2) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920A