Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Smiths Medical ASD Inc. recalls smiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.
Products
  • smiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes
Where it was sold

Worldwide

Nationwide

Units affected
41691 units
About this recall
smiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes, labeled as follows: 1) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 7.0 mm BLUselect trach tube, Item Number 101/561/070; 2) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 8.0 mm BLUselect trach tube, Item Number 101/561/080; 3) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 9.0 mm BLUselect trach tube, Item Number 101/561/090; 4) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 7.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/070; 5) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 8.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/080; 6) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 9.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/090; 7) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 7.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/070; 8) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 8.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/080; 9) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 9.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/090; 10) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 7.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/070; 11) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 8.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/080; 12) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 9.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/090