Lutronic Corporation recalls Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
- Products
-
- Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser
- Company
- Lutronic Corporation
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Canada and European Union.
Nationwide
- Barcodes
08800291820200- Units affected
- 1065
- About this recall
- Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment