Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Lutronic Corporation recalls Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
Products
  • Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada and European Union.

Nationwide

Barcodes
08800291820200
Units affected
1065
About this recall
Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment