Mazor Robotics Ltd recalls Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.
- Products
-
- Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument
- Company
- Mazor Robotics Ltd
- Where it was sold
-
US Nationwide distribution in PR.
Nationwide
- Barcodes
07290109180465072901091811580729010918409807290115751376- Units affected
- 94 devices
- About this recall
- Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument