Smith & Nephew, Inc. recalls DIAMOND POINT

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Smith & Nephew, Inc., has initiated a field action to voluntarily remove two batches of K-Wires due to a packaging error. A complaint was received indicating that a package contained K-WIRE .062X9 2PT DM wires instead of K-WIRE .045X9 2PT DM as described on the product label.
Products
  • DIAMOND POINT
Where it was sold

US, Canada, Norway

Units affected
24 units
About this recall
DIAMOND POINT, BOTH ENDS K-WIRE, REF 128042, metallic bone fixation fastener