GE Healthcare recalls Core Services Hub V8
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
- Products
-
- Core Services Hub V8
- Company
- GE Healthcare
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Units affected
- 250 systems
- About this recall
- Core Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC) services that enables enterprise imaging workflows, All versions