Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

B Braun Medical Inc recalls STREAMLINE BLOODLINE LONG VERSION

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Products
  • STREAMLINE BLOODLINE LONG VERSION
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
04046955348909
Units affected
2021376 units
About this recall
STREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095L