Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Healthcare recalls GE Healthcare Centricty PACS

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
Products
  • GE Healthcare Centricty PACS
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
008406821244300084068210480700840682145572
Units affected
2138 units
About this recall
GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.