Biomerieux Inc recalls VITEK 2 AST-N461 TEST KIT
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
- Products
-
- VITEK 2 AST-N461 TEST KIT
- Company
- Biomerieux Inc
- Where it was sold
-
International distribution to the countries of Austria, Germany, Luxembourg.
- Barcodes
03573026637712- Units affected
- 97680 units
- About this recall
- VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811