Philips North America Llc recalls MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Products
-
- MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120
- Company
- Philips North America Llc
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;
Nationwide
- Barcodes
884838104112884838112858884838104402- Units affected
- 37
- About this recall
- MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);