Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America recalls IntelliVue Patient Monitor MX400 (Model Number 866060)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
Products
  • IntelliVue Patient Monitor MX400 (Model Number 866060)
Where it was sold

US Nationwide. Global Distribution.

Nationwide

Barcodes
0088483803875200884838057562008848380387690088483803877600884838038783
Units affected
545 devices
About this recall
IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)