Philips North America recalls IntelliVue Patient Monitor MX400 (Model Number 866060)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
- Products
-
- IntelliVue Patient Monitor MX400 (Model Number 866060)
- Company
- Philips North America
- Where it was sold
-
US Nationwide. Global Distribution.
Nationwide
- Barcodes
0088483803875200884838057562008848380387690088483803877600884838038783- Units affected
- 545 devices
- About this recall
- IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)