Tandem Diabetes Care, Inc. recalls Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
- Products
-
- Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application
- Company
- Tandem Diabetes Care, Inc.
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
0038915247074000389152470641- Units affected
- 177
- About this recall
- Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No