Ambu Inc. recalls
Ambu Inc. is named on 4 recalls in Recall Relay's records since November 2023, from FDA; the most recent was July 9, 2025.
All 4, newest first.
-
Class Iofficial
Ambu Inc. recalls Ambu SPUR II Labeled as the following: 1 SPUR II Adult ResuscitatorPotential for the manometer port being blocked rendering the manometer non-functional.
-
Class IIofficial
Ambu Inc. recalls Ambu aScope" 5 Broncho HD 5.6/2.8 Sampler Set. Model Number: 622002000USIncorrect labeling in which the front red pouch label did not match the actual size of the medical device. The label on the front of the scope incorrectly states 5.0/2.2 while the back label, shipping box and product inside correctly states 5.6/2.8.
-
Class IIofficial
Ambu Inc. recalls Ambu¿ VivaSight 2 DLT Endobronchial tubeAmbu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.
-
Class IIofficial
Ambu Inc. recalls Ambu aView 2 A MonitorInstructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Ambu Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts