Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ambu Inc. recalls Ambu aView 2 A Monitor

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.
Products
  • Ambu aView 2 A Monitor
Company
Ambu Inc.
Where it was sold

Domestic distribution nationwide.

Nationwide

Barcodes
57074801450895707480156757
Units affected
11845 distributed US
About this recall
Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2