Angiodynamics, Inc. recalls
Angiodynamics, Inc. is named on 8 recalls in Recall Relay's records since November 2023, from FDA (56 recalled products: FDA lists each product as its own recall); the most recent was March 3, 2026.
All 56 recalled products, newest first.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft -Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls AccuVu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls AccuVu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls AccuVu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls AccuVu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls Soft-Vu Angiographic CatheterThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
-
Class IIofficial
Angiodynamics, Inc. recalls IsoLoc Short Profile Prostate Immobilization Treatment Balloon Device Kit; UPN/Catalog Number: ISO-60;Product is mislabeled with the incorrect fill volume.
-
Class IIofficial
Angiodynamics, Inc. recalls AURYON LASER SYSTEM 100-120 VAC. Model Number: EXM001. Product Number: EXM-2001-1100Potential for procedural delays or interruptions during use of the Auryon Atherectomy System due to the Auryon Atherectomy System not advancing to the Activation (Ready) mode.
-
Class IIofficial
Angiodynamics, Inc. recalls SOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length)
-
Class IIofficial
Angiodynamics, Inc. recalls SOFT-VU KUMPE 5F X 40CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732702Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length)
-
Class IIofficial
Angiodynamics, Inc. recalls IsoLoc Prostate Immobilization Treatment Balloon Device KitA single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 28 CM STRAIGHT VASC-PAKXXX
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 24CM STR. BASIC KITTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAFLOW 2Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAFLOW 2Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 28CM STR. BASIC KITTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAFLOW 2Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 32 CM STRAIGHT VASC-PAKTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 28 CM STRAIGHT VASC-PAKTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls BioFlo DuraMax 15.5F 24cm Single Valve Sheath VascPak Kit- Indicated for use in attaining Long-Term vascular aTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAFLOW 2Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAFLOW 2Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 28CM STR. BASIC KITTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 24CM STR. BASIC KITTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAMAX STACKED TIP 32CM STR. BASIC KITTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls DURAFLOW 2Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 2.75" PG -Used for the percutaneous introduction of a guidewireNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guideNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewireNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewiNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 4F X 10CM STIFF .018 NI/TU ECHO 1.6" PG- Used for the percutaneous introduction of a guidewireNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" PG-Used for the percutaneous introduction of a guidewireNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STIFF .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewireNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 4F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewireNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 4F X 10 CM STD .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire iNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire iNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire iNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire iNon-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
-
Class IIofficial
Angiodynamics, Inc. recalls PulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20Rotating adaptor O-ring manufacturing defect impacting seal integrity which may lead to a leak during infusion of medication.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.